Testing Equipments

Thursday, 4 April 2019

Gradation of Coarse Aggregates

Objective : To determination of particle size distribution of coarse aggregates by sieving or
screening.

Reference : IS : 2386 ( Part I) – 1963, IS: 383-1970, IS : 460-1962

Theory :
Grading refers to the determination of the particle-size distribution for aggregate. Grading limits and
maximum aggregate size are specified because grading and size affect the amount of aggregate used as well as cement and water requirements, workability, pumpability, and durability of concrete. In general, if the water-cement ratio is chosen correctly, a wide range in grading can be used without a major effect on strength. When gap-graded aggregate are specified, certain particle sizes of aggregate are omitted from the size continuum. Gap-graded aggregate are used to obtain uniform textures in exposed aggregate concrete. Close control of mix proportions is necessary to avoid segregation.

Apparatus :
Test Sieves conforming to IS : 460-1962 Specification of 80 mm, 40 mm, 20 mm, 10 mm, 4.75 mm, Balance, Gauging Trowel, Stop Watch, etc.

Procedure :
1. The sample shall be brought to an air-dry condition before weighing and sieving. This may be achieved either by drying at room temperature or by heating at a temperature of 100‖ to 110°C. The air-dry sample shall be weighed and sieved successively on the appropriate sieves starting with the largest. Care shall be taken to ensure that the sieves are clean before use.

2. Each sieve shall be shaken separately over a clean tray until not more than a trace passes, but in any case for a period of not less than two minutes. The shaking shall be done with a varied motion, backward sand forwards, left to right, circular clockwise and anti-clockwise, and with frequent jarring, so that the material is kept moving over the sieve surface in frequently changing directions.

3. Material shall not be forced through the sieve by hand pressure. Lumps of fine material, if present, may be broken by gentle pressure with fingers against the side of the sieve.

4. On completion of sieving, the material retained on each sieve, together with any material cleaned from the mesh, shall be weighed.

Observation :



Conclusion / Result :
The results may be observed with the below table

No comments:

Post a Comment